  Datasheet File OCR Text:
  switch mode power supply medical adapter AMD-24V060WXYA 1/5 rev: 00.2, 3/18/2011 summary the AMD-24V060WXYA is the medical adapter from one of the world?s largest power supply manufacturers -delta. the product offers a regulated output voltage of 24v, along with wide temperature range from 0c to +40c and a minimum holdup time of 15ms. the state-of-the-art design is made to withstand medical equi pment requirements. the rugged, ultra-compac t plastic case with shock and vibration resistants. the 60 watts medical power supply provides overvoltage, overload and over temperature protection. due to the wide input voltage range from 90 to 264vac (nominal input: 100vac to 240 vac), the delta?s medical power supply is worldwide usable. the input socket & output connector are made for easy installation. features: ? 60watts output power. ? 100-240vac universal input. ? ac receptacle with c8 & c14 option (iec320ac receptacle) ? efficiency > 87% ? no load power consumption < 0.5w ? energy star compliant (level v) ? low leakage current ? brownout & brownout recovery ? medical safety standard compliant ul60601- 11/en60601-1/csa60601-1 ? mtbf: > 600,000 hrs. model: amd-24v060wx y a
switch mode power supply medical adapter AMD-24V060WXYA 2/5 rev: 00.2, 3/18/2011 technical data 60w medical adapter input data (1) nominal input voltage (wide-range input) 100v-240vac input voltage range 90v-264vac frequency 47-63 hz max input current 1.2a max @ 115 vac, 0.7a max @ 230 vac. inrush current limitation. i 2 t (+25 c) typ. 40a @ 115vac & 80a @ 230vac efficiency (at 115vac and nominal values) >87% no load power consumption >0.5w mains buffering at nominal load (typ.) 15 msec min @ 115 vac turn-on time after applying the mains voltage 3sec transient surge voltage protection varistor input fuse, internal (device protection) t 3.15 ah / 250v discharge current to pe < 0.25ma connection method pluggable output data (2) nominal output voltage u n / tolerance 24v 2% setting range of the output voltage 23.5v-24.5v nominal output current in with convection cooling:-20 to +50c 2.5a max. power dissipation idling/nominal load approx. 60w ripple/ noise (20 mhz) <240mvpp @ +25 c line regulation + 0.5% load regulation : exclude output cord : + 1% include output cord + 5% protection : over current (auto recovery) over voltage (latch) over temp (latch) certification/standards electrical safety (of information technology equipment) ur/cur as per std ul60601-1/csa60601-1, cb scheme to en/iec60601-1, nemko en60601-1 ce mark lvd and emc rohs compliant yes
switch mode power supply medical adapter AMD-24V060WXYA 3/5 rev: 00.2, 3/18/2011 en55011 corresponds to cispr11 / en55022 corresponds to cispr22 / em 61000 corresponds to iec 1000 1) criterion a: normal operating behavior within the defined limits. 2) criterion b: temporary impairment to operational behavior that is corrected by the device itself. 3) symmetrical: conductor to conductor. 4) asymmetrical: conductor to ground. 5) class b: area of application industry and residential general data mtbf > 600,000 hrs type of housing plastic enclosed dimensions (w / h / d) + mounting rail 61w * 39.5h * 120l mm weight 0.28kg. climatic data ambient temperature (operating) ambient temperature (storage) 0 c to 40 c -35 c to + 85 c humidity 5% to 95% non-condensing in conformance with emc guideline 89/336/eec and low voltage directive 73/23/eec emc (electromagnetic compatibility) fcc (title 47 part 15 class b), en55022 (cispr 22 class b) immunity to interference according to en 61000-6-1 ? en 61000-4-2 2) housing discharge of static electricity (esd) contact discharge: air discharge : level 4 8 kv 15 kv ? en 61000-4-3 1) housing electromagnetic hf field frequency/field intensity: level 3 80hz ? 1ghz / 10v/m, with 1khz tone/80% modulation ? en 61000-4-4 2) input fast transients (burst): l,n,pe : signal: level 3 2kv 3) - ? en 61000-4-5 2) input : surge voltage capacities (surge): common mode : differential mode : level 3 2kv 3) 1kv 4) ? en 61000-4-6 1) i/o: ? conducted disturbance frequency / u o : level 3 0.15mhz - 80mhz / 10vrms. ? en 61000-4-11 2) voltage dips self recoverable, no damage noise emission according to en 61000-6-3 ? emitted radio interface ? radio interference voltage en55011 (en55022) class b 5) en55011 (en55022) class b 5)
switch mode power supply medical adapter AMD-24V060WXYA 4/5 rev: 00.2, 3/18/2011 numbering for medical adapter for example: AMD-24V060WXYA xx x xxxx xxxx x xx xx???????.product type am medical adapter xx x xxxx xxxx x xx x???????.model d desktop xx x xxxx xxxx x xx xx xx???????.voltage xx x xxxx xxxx x xx xxxx?????watt 030w 30 watt 060w 60 watt 080w 80 watt xx x xxxx xxxx x xx x?????...input configuration 2 2 pin ac 3 3 pin ac xx x xxxx xxxx x x x x?????...connector type a inner diameter 2,5 mm b inner diameter 2,1 mm *** outer di ameter 5,5 mm. xx x xxxx xxxx xx x x???????.variation (eg, customer, material etc.) a delta standard - product 24v 24 voltage
switch mode power supply medical adapter AMD-24V060WXYA 5/5 rev: 00.2, 3/18/2011 mechanical drawing


▲Up To Search▲   

 
Price & Availability of AMD-24V060WXYA

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X